Leave Your Message
AWA6161A (IEC126)Type Acoustic Coupler
Artificial Ear
From Sound Level Meters and Acoustic Imagers to Calibrators and sound level sensor for professional testing and research.
Inquiry Now

AWA6161A (IEC126)Type Acoustic Coupler

Standard-Compliant Design — Manufactured per SJ/Z9144-1987, equivalent to IEC60318-5 (IEC60126), ensuring internationally recognized measurement accuracy.

High-Precision Frequency Response — Delivers ≤±0.5 dB accuracy across 200–5000 Hz and ≤±1 dB across the full 125–8000 Hz range.

Human Ear Simulation — Accurately replicates the acoustic behavior of the human ear canal for realistic in-ear headphone testing.

Flexible Microphone Compatibility — Works with both φ12.7 mm (1/2") and φ23.77 mm (1") SPL condenser microphones, offering setup flexibility.

Complete System Integration — Pairs seamlessly with AWA14422 microphones and AWA14601B/AWA14604 preamplifiers for a ready-to-use measurement chain.

Also Known as IEC126 Ear — Widely recognized industry-standard equivalent, ensuring cross-compatibility with existing IEC126-based test protocols.

    • prod_icon_03
      Class 1SLM
    • prod_icon_05
      Mobile App
    • prod_icon_07
      IEC61672
    AWA6161A Acoustic Coupler
    01

    Product Description

    The AWA6161A simulated ear is an acoustic coupler that mimics the acoustic characteristics of the human ear. It can be used with a φ12.7mm (1/2 inch) sound pressure level (SPL) capacitive testing microphone (AWA14422 type) and preamplifier (AWA14601B or AWA14604 type), or with a φ23.77mm (1 inch) SPL capacitive testing microphone. It is used with accompanying audio measurement and analysis instruments to measure the electroacoustic characteristics of in-ear headphones. The AWA6161A simulated ear is designed and manufactured according to SJ/Z9144-1987 "Measurement of Couplers for Hearing Aids with In-Ear Headphones and In-Ear Headphones," which is equivalent to IEC60318-5 (IEC60126) standard. Therefore, this simulated ear is also known as the IEC60126 simulated ear or IEC126 ear. The photo shown in Figure 1 shows the assembly of the AWA6161A simulator ear, AWA14601B preamplifier, AWA14422 test microphone, and 1-inch to 1/2-inch adapter.

    Product Video

    Main Performance Indicators

    AWA6161A (IEC126)Type Acoustic Coupler

    Frequency range

     

    125~8000Hz

    Frequency response accuracy

    ≤±0.5dB(200Hz~5000Hz),≤±1dB(<125Hz~8000Hz)

    Standards compliance

    IEC60318-5(IEC60126)

    Aihua Instruments TO KNOW MORE ABOUT Aihua Instruments, PLEASE CONTACT US!

    Our experts will solve them in no time.